Updated programs
| Program Number | Gel Types | Program name | Default Run Time (min.) | Maximal Run Time (min.) |
|---|---|---|---|---|
| 0 | Pre-run | PRE-RUN | 2 | 2 |
| 1 | E-Gel® (0.8%, 1.2%, 2%) | E-Gel 0.8-2.0% | 26 | 40 |
| 2 | E-Gel® 4% | E-Gel 4% | 30 | 40 |
| 3 | E-Gel® double comb (0.8%, 2%) | E-Gel DC | 13 | 20 |
| 4 | E-Gel® CloneWell™ 0.8% | CloneWell 0.8% | 12 | 60 |
| 5 | E-Gel® CloneWell™ Reverse run | REVERSE E-Gel | 2 | 3 |
| 6 | E-Gel® (0.8%, 1.2%, 2%) | SPEED E-Gel | 7 | 7 |
| 7 | E-Gel® EX 1%, 2% | E-Gel EX 1-2% | 10 | 20 |
| 8 | E-Gel® EX 4% | E-Gel EX 4% | 15 | 20 |
| 9 | SizeSelect™ 2% | SizeSelect 2% | 8 | 20 |
System requirements:
- Supported Operating Systems: Windows XP and Windows Vista
- PC with 600 MHz equivalent or higher processor
- 128 MB of hard drive space
Additional requirements:

- Cable: USB to USB Mini Cable.
- Available for purchase from Invitrogen. Request catalog number G6300.
Instructions
- Disconnect the electrical plug of the iBase from the electrical outlet.
- Please make sure the USB cable is not connected.
- Press and hold the "Go” button (red button).
- Continue holding the "Go" button and Insert the power plug into the electrical outlet, and then connect the cable to the iBase unit.
- Release the "Go" button and connect the iBase to this computer with a USB A to B cable (A end into the computer, B end into the iBase). The computer should now begin to search for the iBase. This step may take several minutes.
- The program will inform it is searching for the iBase.
- The program will inform the iBase has been found.
- Please press ‘Next’ to begin the iBase Firmware Update.
- Please do not disconnect or use the device until iBase Updater is complete !
- The update was successful.
- Please disconnect the USB cable from the iBase device.
- The iBase is now updated.
Troubleshooting
- If the update was unsuccessful, the message "The Update failed" will appear.
- Retry the program and if the problem persists contact Technical Support for further assistance.