Your instrument may not be adequately calibrated

Instrument calibration

To ensure optimal performance of your thermal cycler or sequence detection system, Applied Biosystems strongly recommends that you regularly calibrate the system. Calibration schedules vary by instrument; refer to the table below for the types of calibration and appropriate intervals.

Instrument calibration intervals

Calibration Type Instrument Interval
ROI (Spatial) 7300/7500/7500 Fast Real-time PCR Systems StepOne™ / StepOnePlus™ SystemsEvery 6 months or As needed to verify the instrument’s performance
Background 7900HT Fast Real-Time PCR System 7300/7500/7500 Fast Real-time PCR Systems Every month or Before performing a pure dye calibration or When installing an uncalibrated block
Pure dye spectra 7900HT Fast Real-Time PCR System 7300/7500/7500 Fast Real-time PCR Systems Every 6 months or When installing an uncalibrated block
Instrument verification 7900HT Fast Real-Time PCR System 7300/7500/7500 Fast Real-time PCR Systems Every 6 months or As needed to verify the instrument’s performance