Physical Characteristics | | - Instrument type: Benchtop cell analyzer and suspension cell-based assay platform
- Instrument dimensions: 11 ½” (W) X 11 ½” (H) X 17 ½” (D)
- Weight: 19.4 lbs (8.8 kg)
- Operating power: 100–240 VAC, 2.5 A, 120 V
- Electrical input: 12 VDC, 13 A
- Installation site: Indoor use only
- Operating temperature: 15–29°C
- Operating humidity: < 80% (non-condensing)
|
|
| Technical Specifications | | - Processing time: 10 seconds to 2 minutes, depending on the number of fields captured and complexity of assay selected
- Sample concentration range: 1 × 105–1 × 107 cells/mL
- Recommended sample diameter range: 5–60 μm
- Required sample volume: 25 μL
- Firmware: Tali™ Image Cytometer Firmware (visit www.invitrogen.com/tali for updates)
|
|
| Optics | | - Optics: 3 channels (bright field, green fluorescence, red fluorescence)
- Excitation: Green channel LED: 458 ± 20 nm
- Red channel LED: 530 ± 20 nm
- Filters: Green channel: 466/40 EX, 495 LP Di, 525/50 EM
- Red channel: 543/22 EX, 580 LP Di, 585 LP EM
- Camera: 1.3 Mega pixels, 4X objective, 4X or 16X digital zoom
|
|
| Tali™ Cellular Analysis Slide | | - Material: Polylactic acid (PLA)
- Dimensions: 110 mm (w) × 24 mm (d) × 1.9 mm (h)
|